  1. Cup resists attachment and detachment from Cup Holder.

  2. Spindle Housing tracks poorly.

  3. Bob does not rotate when running a test.

  4. Power switch indicator or LEDs do not light up when the power is turned on (and unit is properly plugged in).

  5. The viscometer is making a high pitch noise.



Cup resists attachment and detachment from Cup Holder.

Likely Cause:  Dried sample material in and around the Cup Holder.
Solution:  Clean thoroughly with steel wool.

Likely Cause:  Bur has formed on Cup's camlock groove.
Solution:  File bur off.

Like Cause:  Cup is bent.
Solution:  Replace with a new cup immediately.




Sprindle Housing tracks poorly.

Likely Cause:  Sample material on Slide.
Solution:  Clean thoroughly with stiff brush.

Likely Cause:  A bur has formed on the Slide.
Solution:  Use fine sandpaper or emery cloth to remove bur and lubmircate.




 
 Bob does not rotate when running a test.
 

Likely Cause: DC fuse (located in back of unit) has blown.
Solution: Replace with AGC 20 fuse only.



>> Return to top <<

Power switch indicator or LEDs do not light up when the power is turned on (and unit is properly plugged in).

Likely Cause: AC fuse is blown.
Solution: Contact Kaltec's Service Department before replacing.


>> Return to top <<



The viscometer is making a high pitch noise.

Likely Cause: DC fuse (located in back of unit) has blown.
Solution: Replace with AGC 20 fuse only.